116 8000 Operator’s Manual
5 Maintenance and Troubleshooting
6 Enter the following parameters:
7 Press ESC until the Ready screen displays.
8 Press START PROG.
9 Enter the program number for this cleaning program and press
ENTER.
10 Press 1 > 1 > 1 > 1 to start the cleaning program.
After the 8000 pulls three rows of samples for cleaning purposes,
complete the following steps to ensure the needles are completely
purged of cleaning solution:
1 From the Ready screen, press MENU > 7 > 4 > R.
2 Press ESC. The Manual Operations screen displays.
3 Select option 3, Open/Close Valves, to open the valves.
4 Select option 2, Lower Valves, to lower the needles.
5 Select option 1, Lift Valves, to raise the needles.
Table 3 0 Program Variables
Selection Response
2 Sample Time Points Row 0: Enter 000:05 and press ENTER.
Row 1: Enter 000:10 and press ENTER.
Row 2: Enter 000:15 and press ENTER.
Row 3: Enter 000:00 and press ENTER.
3 Sample Volume Enter a sample volume based on the size vial used in the 8000 sample tray and
press ENTER. Ensure Rows 0 - 2 in the sample tray contain empty vials.
4 Set Headers Enter the requested information as applicable. Press ESC to return to the
Program Variables screen.
5 Prime Volume Enter 10 and press ENTER.
6 Purge Volume Enter 15 and press ENTER.
7 Filter Chg This option is only applicable if your system configuration includes a filter
changer.
Set the number of samples per filter to 1 and press ENTER.
Zobrazit stránku 115