Setting Up the 8000 3
8000 Operator’s Manual 35
• If your system configuration includes the syringe pump, the power
receptacle, and power switch are located on the left side of the unit.
Connect the power cord and turn on the syringe pump.
• If your system configuration includes the filter changer, the power
receptacle, and power switch are located on the rear of the unit.
Connect the power cord and turn on the filter changer.
The 8000 has a dual voltage (115 V / 230 V) option on the power entry
module. Ensure the voltage indicated on the fuse drawer matches the
power outlet. Connect the power cord to the rear of the 8000. Turn on
the 8000. The system monitor illuminates. The system monitor displays
the initial status screen and then changes to the Ready screen.
The firmware version (program revision) displays. This is the only time
the firmware version displays. Record the number below and refer to it
if you need to call the Dissolution Systems Service Department.
While the unit is powering up, the dispensing arm returns to its home
position at the rear of the unit and the firmware program initializes. If
the dispensing arm is already at the home position when the unit is
powered up, the initial status screen displays for three seconds only.
When the initial status screen disappears, the Ready screen displays.
From the Ready screen, you can access the Main Menu or any of the
manual functions.
8000 FRACTION COLLECTOR
PROGRAM REVISION x.xx
INITIALIZING....PLEASE WAIT
TIME: 01/01/12 21:56:40 ELAPSE: 000:00:00
Firmware version
READY
PRG# 01 R-M: ENABLED VOLUME: 5 mL
TIME: 01/01/12 21:56:40 ELAPSE: 000:00:00