106 8000 Operator’s Manual
5 Maintenance and Troubleshooting
Preventive Maintenance
Perform the following maintenance when necessary:
Clean System with Peristaltic Pump
Use the Clean System function often to keep the valves, needles, and
sample lines clean. The timing of the valves opening and closing with
the pump rotations creates cavitation in the valves. Cleaning the system
dislodges any particulate matter that may affect the operation of the
system. To prolong proper functionality of the 8000 system, it is
recommended to perform a cleaning cycle after each dissolution test
program.
Apparatus 1 and 2
1 Place the rinse tray on the vessel table of the dissolution apparatus.
2 Fill the rinse tray with an appropriate cleaning solution, usually
water.
3 Fill the rinse chamber of the media rinse reservoir with an
appropriate cleaning solution, usually water.
4 Lower the sampling cannulas. See the appropriate dissolution
apparatus operators manual for information on how to operate the
sampling cannulas.
WARNING
The 8000 and accessories contain electrical circuits, devices, and
components operating at dangerous voltages. Contact with these
circuits, devices, and components can cause death, serious
injury, or painful electric shock.
NOTE
Prior to beginning the clean system function, verify an appropriate
manual prime time and purge time have been entered via the Main
Menu.