2-102 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
3GPP W-CDMA Base Station Performance Testing Option 200
Part A. Modulated Interferer Measurements
1. Introduce the modulated negative frequency offset out-of-band interferer.
The out-of-band interfering stimulus is also made available at the RF output of ESG 2.
In accordance with the 3GPP W-CDMA RF specification, the center frequency of this
interferer must be offset a minimum of 10 MHz below the center frequency of
fundamental W-CDMA signal. The amplitude of this interferer is
40.00 dBm.
On ESG 2:
Press
Frequency and enter a value between 1900 MHz and 1920 MHz. This value must
be at least 10 MHz below the center frequency of the fundamental W-CDMA signal.
2. Measure the BER of the BS in the presence of the modulated negative frequency offset
out-of-band interferer.
The controller or BER test platform is measuring the bit error ratio of the BS in the
presence of both the fundamental W-CDMA signal from ESG 1 and the modulated
negative frequency offset out-of-band interferer from ESG 2.
3. Record the BER measurement results for the modulated negative frequency offset
out-of-band interferer in the column titled, “Negative Frequency Offset Interferer BER
Value” in Table 2-54.
4. Introduce the modulated positive frequency offset out-of-band interferer.
The out-of-band interfering stimulus is also made available at the RF output of ESG 2.
In accordance with the 3GPP W-CDMA RF specification, the center frequency of this
interferer must be offset a minimum of 10 MHz above the center frequency of
fundamental W-CDMA signal. The amplitude of this interferer is
40.00 dBm.
On ESG 2:
Press
Frequency and enter a value between 1980 MHz and 2000 MHz. This value must
be at least 10 MHz above the center frequency of the fundamental W-CDMA signal.
5. Measure the BER of the BS in the presence of the modulated positive frequency offset
out-of-band interferer.
The controller or BER test platform is measuring the bit error ratio of the BS in the
presence of both the fundamental W-CDMA signal from ESG 1 and the modulated
positive frequency offset out-of-band interferer from ESG 2.
6. Record the BER measurement results for the modulated positive frequency offset
out-of-band interferer in the column titled, “Positive Frequency Offset Interferer BER
Value” in Table 2-54.
Part B. CW Interferer Measurements
NOTE An external signal source capable of providing CW at an amplitude of
15 dBm across a frequency range of 1.000 MHz to 12.750 GHz is required for
out-of-band performance testing.
Zobrazit stránku 117
1 2 ... 113 114 115 116 117 118 119 120 121 122 123 ... 301 302

Komentáře k této Příručce

Žádné komentáře