3-26 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Softkey Reference ESG Family Signal Generators
E Option 200
Err Insert
This field allows you to select an error insertion mode to be applied to
the data for the transport channel. This field is on the second page of
the table editor. The selections are BLER (block error rate), BER (bit
error rate), and NONE.
Default: NONE
Data Field Location (Uplink):
Mode > CDMA > W-CDMA > Link Up >
Link Control > 2 > Config Transport > Transport Channel # 1 to 6 >
TrCH Setup > Err Insert
Error Bits
This field displays the number of actual bit errors being generated. It is
calculated from the desired BER value as set in the BER field after the
Apply Channel Setup key is pressed and the error bits are inserted into
the data for the transport channel. To insert the error bits, the Err
Insert field must be set to BER. This data field is on the second page of
the table editor.
Data Field Location (Uplink):
Mode > CDMA > W-CDMA > Link Up >
Link Control > 2 > Config Transport > Transport Channel # 1 to 6 >
TrCH Setup > Error Bit
Error Blocks
This field displays the number of the actual error blocks being
generated. It is calculated from the desired BLER value as set in the
BLER field after the
Apply Channel Setup key is pressed and the error
blocks are inserted into the data of the transport channel. To insert the
error blocks, the Err Insert field must be set to BLER. This data field is
on the second page of the table editor.
Data Field Location (Uplink):
Mode > CDMA > W-CDMA > Link Up >
Link Control > 2 > Config Transport > Transport Channel # 1 to 6 >
TrCH Setup > Error Block
Ext Clock Polarity Neg Pos
This key selects the external clock polarity to either the positive (Pos)
edge or negative (
Neg) edge. This is valid only when BBG Data Clock is
set to
Ext.
Default: Positive
Softkey Location (Uplink):
Mode > CDMA > W-CDMA > Link Up >
BBG Data Clock Setup > Ext Clock Polarity Neg Pos