Agilent Technologies ESG Specifikace Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 302
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
2-6 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
Two ESG Downlink Measurement Setup Option 200
Two ESG Downlink Measurement Setup
This setup provides the capability to test 3GPP receiver designs for capability to test
adjacent channel selectivity, blocking, intermodulation, maximum input, receiver
sensitivity, and spurious response.
The following example configures a 3GPP receiver test stimulus.
Required Equipment
two ESG-D Family Signal Generators with Option 200 and UN8
one E4406A VSA Series Transmitter Tester with Option BAF
one function generator (with an external frequency reference input or a 10 MHz
frequency reference output) capable of providing an external ESG data clock input with
the following characteristics:
wave type: square
amplitude: 2.5 V
rms
frequency: 3.84 MHz
DC offset: +1.25 V
13 BNC cables (long, equal lengths)
6 BNC cables (short, equal lengths)
6 BNC tee splitters
3 BNC to BNC connectors
3 Type-N to BNC adapters (for ESG and VSA RF Output)
System Configuration Process
The two-ESG system configuration process is divided into the following subsections:
Connect the Equipment” on page 2-7
The W-CDMA Configuration” on page 2-8
Configure ESG 1” on page 2-10
Configure ESG 2” on page 2-11
Adjust the Relative Power Mapping” on page 2-14
Generate the Signal at the RF Output” on page 2-16
Configure the E4406A Option BAF VSA Series Transmitter Tester” on page 2-16
Zobrazit stránku 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 301 302

Komentáře k této Příručce