Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide 2-115
ESG Family Signal Generators Using Functions
Option 200 Troubleshooting
Troubleshooting
Use the following checklists to help troubleshoot problems with Option 200.
NOTE In some cases, the ESG may allow invalid channel coding settings without
reporting an error. If you experience problems, ensure that the setting of the
ESG(s) conform to the 3GPP Technical Specification.
Tips and Tricks
There is a Tips and Tricks section located on the Agilent Technologies ESG support web
site that may provide you assistance in the operation of the Option 200 firmware. Perform
the following steps to reach the appropriate web site:
1. Enter and go to www.agilent.com/find/esg.
2. Select your model of ESG under the Products list.
3. Select Software Patches and Drivers, in the next window, under the Additional
Information list.
4. Select the firmware version currently loaded on your ESG, shown under the Software
Updates section.
The firmware version is displayed when the ESG is powered on or it can be viewed by
pressing the following keys and viewing the listed field:
Utility > Instrument Info/Help Mode > Diagnostic Info > Firmware Revision
5. Locate the Tips and Tricks section and follow any instructions that may be appropriate
to your situation.
Can’t Turn on W-CDMA
❏ At least one W-CDMA channel must be on.
Can’t See Channels Displayed on Transmitter Tester
Ensure that the following are true:
❏ W-CDMA channels are on.
❏ W-CDMA is on (W-CDMA and I/Q annunciators are displayed). I/Q must be toggled on to
modulate the RF output.
❏ ESG RF output is on (RF ON annunciator is displayed).
❏ ESG RF is modulated (MOD ON annunciator is displayed).
❏ Scrambling Code (BS Setup) of ESG matches primary scramble code of VSA.
❏ For two-ESG system configurations, the I/Q Source is set to Ext I/Q and BBG Data
Clock set to Ext.