Fully Coded 3GPP W-CDMA Personality Users and Programming Guide 2-33
ESG Family Signal Generators Using Functions
Option 200 3GPP W-CDMA User Equipment Performance Tests
At this point, the controller must configure the UE into a test mode. With the UE
operating in test mode, the controller assigns the UE a dedicated physical channel on
which to receive and demodulate the transmitted data.
3. Measure the Bit Error Ratio (BER) of the data received by the UE on RMC 12.2 kbps
(25.101 V3.3).
The controller (or BER test platform) is measuring the bit error ratio of the UE exposed
to Î
or
(the RF output of ESG 1) at an arbitrary amplitude. This process verifies that the
BER of the UE is within specification prior to increasing the amplitude of Î
or
.
4. In Table 2-17, record the measured BER value in the column titled, “Verification BER
Value.”
5. Compare the measured BER value in the column titled, “Verification BER Value” with
the limits from the column titled, “Specification Limitations.”
If the measured value is within the limits, continue to subsection 2.
If the measured value is outside the limits, determine the cause of failure and repair
the UE. After repair, return to Step 1 of this subsection.
Subsection 2. Maximum Input Level Measurement
Complete the following steps to perform a maximum input level measurement on your UE.
1. Increase the amplitude of Î
or
to the required value.
In accordance with the 3GPP W-CDMA RF specification, increase the amplitude of Î
or
to
25.0 dBm.
On ESG 1:
Press
Amplitude >
25
> dBm.
2. Measure the BER of the data received by the UE on RMC 12.2 kbps (25.101 V3.3) at the
increased amplitude.
The controller (or BER test platform) is measuring the bit error ratio of the UE with an
Î
or
amplitude of
25.0 dBm.
3. In Table 2-17, record the measured BER value in the column titled, “Maximum
Amplitude BER Value.”
4. Compare the recorded BER value in the column titled, “Maximum Amplitude BER
Value” to the allowable values in the column titled, “Specification Limitations” and
record the status of the UE as PASS or FAIL in the column titled, “PASS/FAIL Status.”