2-40 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
3GPP W-CDMA User Equipment Performance Tests Option 200
At this point, the controller must configure the UE into a test mode. With the UE
operating in test mode, the controller assigns the UE a dedicated physical channel on
which to receive and demodulate the transmitted data.
3. Measure the Bit Error Ratio (BER) of the data received by the UE on RMC 12.2 kbps
(25.101 V3.3).
The controller (or BER test platform) is measuring the bit error ratio of the UE exposed
to Î
or
(the RF output of ESG 1) only. This process verifies that the BER of the UE is
within specification prior to the introduction of the adjacent channel interferer.
4. In Table 2-22, record the measured BER value in the column titled, “Verification BER
Value.”
5. Compare the measured BER value in the column titled, “Verification BER Value” with
the limits from the column titled, “Specification Limitations.”
• If the measured value is within the limits, continue to subsection 2.
• If the measured value is outside the limits, determine the cause of failure and repair
the UE. After repair, return to Step 1 of this subsection.
Subsection 2. ACS Measurements
Complete the following steps to perform adjacent channel selectivity measurements on
your UE.
Figure 2-6 ACS Measurement Equipment Setup
1. Connect the additional equipment to the test system as shown in Figure 2-6.
a. Connect the RF output of ESG 1 to one input of the signal combiner.
b. Connect the RF output of ESG 2 to one input of the signal combiner.
c. Connect the signal combiner output to the RF input of the UE.