Zobrazit stránku 65
2-50 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
3GPP W-CDMA User Equipment Performance Tests Option 200
2. Adjust the amplitude of Î
or
to the required value.
In accordance with the 3GPP specification, adjust the amplitude of Î
or
to
92.7 dBm.
On ESG 1: Press
Amplitude >
92.7
> dBm.
3. Introduce the 10 MHz offset in-band interferer (I
blocking
).
The first in-band interfering stimulus is made available at the RF output of ESG 2. In
accordance with the 3GPP W-CDMA RF specification, the center frequency of this
unwanted interferer (I
blocking
) can be
±
10 MHz from the center frequency of Î
or
. The
amplitude of I
blocking
is
52.0 dBm.
On ESG 2:
Press
RF On/Off (or execute the appropriate SCPI command) until the RF ON
annunciator is displayed.
4. Measure the BER of the data received by the UE on RMC 12.2 kbps (25.101 V3.3) in the
presence of the 10 MHz offset in-band interferer.
The controller (or BER test platform) is measuring the bit error ratio of the UE in the
presence of both RF output waveforms, Î
or
(from ESG 1) and the 10 MHz offset in-band
interferer I
blocking
(from ESG 2).
5. In Table 2-27, record the measured BER value in the column titled, “10 MHz Offset
Interferer BER Value.”
6. Introduce the 15 MHz offset in-band interferer.
The second in-band interfering stimulus is made available at the RF output of ESG 2.
In accordance with the 3GPP W-CDMA RF specification, the center frequency of this
unwanted interferer (I
blocking
) can be
±
15 MHz from the center frequency of Î
or
. The
amplitude of I
blocking
is
44 dBm.
On ESG 2:
a. Press
Frequency and enter a frequency value
±
15 MHz from the center frequency of
Î
or
to change the I
blocking
carrier frequency.
b. Press
Amplitude >
44
> dBm to change the I
blocking
amplitude.
c. Press
RF On/Off until the RF ON annunciator is displayed on ESG 2 to activate the
RF output.
7. Measure the BER of the data received by the UE on RMC 12.2 kbps (25.101 V3.3) in the
presence of the 15 MHz offset in-band interferer.
The controller (or BER test platform) is measuring the bit error ratio of the UE in the
presence of both RF output waveforms, Î
or
(from ESG 1) and the 15 MHz offset in-band
interferer I
blocking
(from ESG 2).
8. In Table 2-27, record the measured BER value in the column titled, “15 MHz Offset
Interferer BER Value.”