Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide 3-49
ESG Family Signal Generators Softkey Reference
Option 200 O
Oversample Ratio
This key sets the oversampling ratio to be applied to the FIR filter
design in the
Define User FIR menu. Although the FIR Values table
editor allows a maximum of 32 symbols, with a maximum oversample
ratio of 32, the selected FIR filter cannot have more than 256
coefficients. Select the number of symbols and the oversample ratio
accordingly.
The oversample ratio value is set at the factory. This value can be
user-defined. The value is not affected by an instrument preset or by a
power cycle.
Range: 1 to 32
Default: 1
Softkey Location (Downlink):
Mode > CDMA > W-CDMA > BS Setup >
Filter >
Edit Item > Define User FIR > Oversample Ratio
Softkey Location (Uplink): Mode > CDMA > W-CDMA > Link Up >
UE Setup > Filter > Edit Item > Define User FIR > Oversample Ratio