Agilent Technologies ESG Specifikace Strana 88

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 302
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 87
2-72 Fully Coded 3GPP W-CDMA Personality User’s and Programming Guide
Using Functions ESG Family Signal Generators
3GPP W-CDMA User Equipment Performance Tests Option 200
Subsection 3. Positive Frequency Offset Interferers IRR Measurements
Complete the following steps to make positive frequency offset measurements on your UE.
1. Introduce the positive frequency offset CW interferer (I
ouw1
).
I
ouw1
is made available at the RF output of the CW source. In accordance with the 3GPP
W-CDMA RF specification, the center frequency of I
ouw1
is set 10 MHz above the center
frequency of Î
or
. The amplitude of I
ouw1
is
46 dBm.
On the CW source:
Set the RF output frequency to 2150 MHz (or 10 MHz above the center frequency of Î
or
).
Set the RF output amplitude to
46 dBm.
Activate the RF output.
2. Introduce the positive frequency offset modulated interferer (I
ouw2
).
I
ouw2
is made available at the RF output of ESG 2. In accordance with the 3GPP
W-CDMA RF specification, the center frequency of I
ouw2
is set 20 MHz above the center
frequency of Î
or
. The amplitude of I
ouw2
is
46 dBm.
On ESG 2:
a. Press
Frequency > 2160 > MHz (or a frequency 20 MHz above the center frequency of
Î
or
) to set the carrier frequency of positive frequency offset modulated interferer
(I
ouw2
).
b. Press
RF On/Off until the RF ON annunciator is displayed on ESG 2 to activate the RF
output.
3. Measure the BER of the data received by the UE on RMC 12.2 kbps (25.101 V3.3) in the
presence of the positive frequency offset CW and modulated interferers (I
ouw1
and
I
ouw2
).
The controller (or BER test platform) is measuring the bit error ratio of the UE in the
presence of all three RF output waveforms, Î
or
(from ESG 1) and the positive frequency
offset interferers I
ouw1
and I
ouw2
(from the CW source and ESG 2, respectively).
4. In Table 2-41, record the measured BER value in the column titled, “Positive Frequency
Offset Interferers BER Value.”
5. Compare the recorded BER value in the column titled, “Positive Frequency Offset
Interferers BER Value” to the allowable values in the column titled, “Specification
Limitations” and record the status of the UE as PASS or FAIL in the column titled,
“PASS/FAIL Status.”
Zobrazit stránku 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 301 302

Komentáře k této Příručce

Žádné komentáře