Fully Coded 3GPP W-CDMA Personality Users and Programming Guide 2-19
ESG Family Signal Generators Using Functions
Option 200 3GPP W-CDMA User Equipment Performance Tests
Required Equipment
The following equipment must be calibrated and must perform within the limits of its
published specifications.
two ESG signal generators configured as a dual-output, two-ESG W-CDMA mobile test
solution
E4406A VSA Series W-CDMA transmitter tester
function generator (with an external frequency reference input) capable of providing an
external ESG data clock input with the following characteristics:
wave type: square
amplitude: 2.5 V
rms
frequency: 3.84 MHz
DC offset: +1.25 V
83712B synthesized CW generator (or another signal source capable of providing CW at
an amplitude range of
46 dBm to
15 dBm with an RF output frequency range of
2.11 GHz to 12.75 GHz)
a UE test mode designed to allow the controller to assign the UE a dedicated physical
channel on which to receive and demodulate the transmitted data
1
A system controller and software capable of providing control data to the UE
software and/or hardware (a BER test platform) capable of measuring, calculating and
storing the UE’s bit error ratio (BER) while exposed to the test stimulus
microwave frequency combiner
BNC cables, tee splitters and adapters
adapters and antennae as required by the UE
NOTE All ESG signal parameters are specified at the front panel RF OUTPUT
connector. Losses due to cabling and signal combining must be considered
when performing the tests.
1. The signal generator does not receive nor act upon uplink data from the UE.