Fully Coded 3GPP W-CDMA Personality Users and Programming Guide 2-55
ESG Family Signal Generators Using Functions
Option 200 3GPP W-CDMA User Equipment Performance Tests
Figure 2-9. Out-of-Band Band 3 Blocking Characteristics Measurements
Equipment Setup 2
5. Measure the BER of the data received by the UE on RMC 12.2 kbps (25.101 V3.3) in the
presence of the positive frequency offset interferer for band 3.
The controller (or BER test platform) is measuring the bit error ratio of the UE in the
presence of both RF output waveforms, Î
or
(from ESG 1) and the band 3 positive
frequency offset interferer I
blocking
(from ESG 2).
6. In Table 2-30, record the measured BER value in the column titled, “Positive Frequency
Offset Interferer BER Value.”
7. Compare both the BER values recorded in columns “Negative Frequency Offset
Interferer BER Value” and “Positive Frequency Offset Interferer BER Value” to the
allowable values in the column titled, “Specification Limitations” (both values must be
within the allowable specification for a PASS status). Record the status of the UE as
PASS or FAIL in the column titled, “PASS/FAIL Status.”