Agilent B1500A/B1505A Programming Guide, Edition 11 2-25
Remote Mode Functions
Measurement Modes
6. After the base hold time, the synchronous source channels force the bias value or
the peak value as follows.
The SMUs start output simultaneously, and then the SPGUs start output in the
order from lower to higher slot number. However the SPGU pulse outputs are
started simultaneously.
The channels keep the output until the end of the sampling measurement.
7. And after the bias hold time, the measurement channels start measurement for
the first sampling point.
If you use multiple measurement channels, the channels perform measurement
in the order defined in the MM command. Parallel measurement is also
available, see “Parallel Measurement Function” on page 2-48.
8. After that, the following operation is repeated with Sampling interval.
Measurement channels start measurement if they are ready to measure.
Measurement channels keep the condition if they are busy.
This operation is repeated until the number of measurement result data reaches
to the specified number of samples.
For the linear sampling with interval < 2 ms, if the total sampling time runs over
Bias hold time + Sampling interval × number, the sampling measurement will be
stopped even if the number of measured data is less than the specified number.
For the log sampling, the B1500A holds only the data that can be plotted on the
log scale in the same distance as close as possible. Only the held data is counted
in the number of measured data.
9. After the sampling measurement, the synchronous source channels change the
output value as follows.
The SPGUs change the output in the order from higher to lower slot number, and
then the SMUs change the output simultaneously. However the SPGU pulse
outputs are stopped simultaneously.
If the SPGU operation mode is not Free Run and the total time of pulse output is
shorter than the sampling measurement time, the SPGU pulse outputs are
stopped simultaneously before the measurement is completed.
SPGU forces the post measurement output value set by the MSP command.
SMU forces the post measurement output value set by the MSC command.
10. The source channel set by the DV or DI command continues the source output.