Agilent Technologies B1500A Servisní příručka Strana 337

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 336
Agilent B1500A/B1505A Programming Guide, Edition 11 4-21
Command Reference
Table 4-4 Voltage Output Ranging Type
NOTE Output ranging
The instrument automatically selects the minimum range that covers the output
value, and applies voltage or current by using the range. For the limited auto
ranging, the instrument does not use the range lower than the specified range value.
For example, if you select the 100 nA limited auto ranging, the instrument never
uses the 10 nA range and below.
range or
vrange
Measurement resource type
Ranging type
HR MP HP MC HC DHC HV
0 Yes Yes Yes Yes Yes Yes Yes Auto ranging
2 Yes Yes Yes 0.2 V limited auto ranging
5 Yes Yes 0.5 V limited auto ranging
20 or 11 Yes Yes Yes Yes Yes Yes 2 V limited auto ranging
50 Yes Yes 5 V limited auto ranging
200 or 12 Yes Yes Yes Yes Yes Yes 20 V limited auto ranging
400 or 13 Yes Yes Yes Yes Yes Yes 40 V limited auto ranging
1000 or 14 Yes Yes Yes 100 V limited auto ranging
2000 or 15 Yes Yes 200 V limited auto ranging
5000 Yes 500 V limited auto ranging
15000 Yes 1500 V limited auto ranging
30000 Yes 3000 V limited auto ranging
range or
vrange
Measurement resource type
Ranging type
UHCU HVMCU UHVU
0 Yes Yes Yes Auto ranging
1000 or 14 Yes 100 V limited auto ranging
15000 Yes 1500 V limited auto ranging
30000 Yes 3000 V limited auto ranging
103 Yes 10 kV limited auto ranging
Zobrazit stránku 336