Agilent Technologies B1500A Servisní příručka Strana 515

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 514
Agilent B1500A/B1505A Programming Guide, Edition 11 4-199
Command Reference
TGSI
TGSI
The TGSI command selects Case 1 or Case 2 effective for the Start Step Output
Setup trigger input set by the TGP port,1,polarity,2 command.
This command is available for the staircase sweep, multi channel sweep, pulsed
spot, pulsed sweep, staircase sweep with pulsed bias, multi channel pulsed spot,
multi channel pulsed sweep, and MFCMU DC/AC/frequency sweep measurements.
Syntax TGSI mode
Parameters mode : Case 1 or Case 2. Integer expression. See Figure 4-5.
1: Case 1 (initial setting).
2: Case 2.
Case 1 waits for a trigger for the first sweep step, and does not wait for a trigger for
the source output after sweep.
Case 2 does not wait for a trigger for the first sweep step, and waits for a trigger for
the source output after sweep.
Example
Statements
OUTPUT @B1500;"TGSI 2"
See Also “TGP” and “TGPC”
Figure 4-5 Trigger Input Example, Staircase Sweep Measurement, Negative Logic
: Measurement
TGP t,1,p,3
TGP t,1,p,2
TGP t,1,p,1
Hold time (Case 2)
or more (Case 1)
Delay time or more
Start Step Measurement
Start Step Output Setup
Start Measurement
Measurement trigger delay
Case 1: With trigger for
first sweep step
Case 2: With trigger for
output after sweep
Case 1
Case 2
Zobrazit stránku 514
1 2 ... 510 511 512 513 514 515 516 517 518 519 520 ... 597 598

Komentáře k této Příručce

Žádné komentáře