4-116 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
*LRN?
PT hold time,pulse width[,pulse period[,trig delay]]
[;PV chnum,output range,base voltage,pulse voltage [,Icomp]] or
[;PI chnum,output range,base current,pulse current [,Vcomp]]
[;PWV chnum,mode,range,base,start,stop,nop[,Icomp]] or
[;PWI chnum,mode,range,base,start,stop,nop[,Vcomp]]<CR/LF^EOI>
37 : Returns the quasi-pulsed source settings:
BDM detection interval[,mode];
BDT hold time,delay time
[;BDV chnum,range,start,stop[,Icomp]]<CR/LF^EOI>
38 : Returns the digital I/O port information:
ERM input pin;ERC2,value <CR/LF^EOI>
40 : Returns channel mapping information:
If multiple channel numbers are translated to another numbers.
ACH actual,program
[;ACH actual,program]....
[;ACH actual,program]<CR/LF^EOI>
If no channel number is defined by the ACH command.
ACH<CR/LF^EOI>
46 : Returns SMU measurement operation mode settings:
CMM chnum,mode[;CMM chnum,mode]....
[;CMM chnum,mode]<CR/LF^EOI>
47 : Returns the sampling measurement settings:
MSC abort,post;MT h_bias,interval,number,h_base;ML mode
[;MV chnum,range,base,bias,comp] or
[;MI chnum,range,base,bias,comp] or
[;MSP chnum[,post [,base]]]
:
<CR/LF^EOI>
49 : Returns the quasi-static CV measurement settings:
QSM abort,post;QSL data,compen;QSZ mode
;QSO mode[,chnum[,Vcomp]]
;QSC mode
;QST cinteg,linteg,hold
,delay1[,delay2]
;QSR range