4-172 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SAR
Syntax SAP chnum,path
Parameters chnum : Channel number of the SMU connected to the ASU. Integer expression.
1 to 10 or 101 to 1001. See Table 4-1 on page 4-16.
path : 0: The ASU output will be connected to the SMU connector side.
1: The ASU output will be connected to the AUX connector side.
Example
Statements
OUTPUT @B1500;"SAP 1,1"
SAR
This function is available for the Agilent B1500 installed with the atto sense and
switch unit (ASU).
Enables or disables the 1 pA range for the auto ranging operation. This command is
effective for the specified channel.
Syntax SAR chnum,mode
Parameters chnum : Channel number of the SMU connected to the ASU. Integer expression.
1 to 10 or 101 to 1001. See Table 4-1 on page 4-16.
mode : 0: Enables 1 pA range for the auto ranging operation.
1: Disables 1 pA range for the auto ranging operation. Initial setting
Example
Statements
OUTPUT @B1500;"SAR 1,0"
SCR
The SCR command scratches the specified program from the internal program
memory.
Syntax SCR [pnum]
Parameters pnum : Program number. Numeric expression. 1 to 2000.
If you do not specify this parameter, this command scratches all
programs stored in the internal program memory.
Example
Statements
OUTPUT @B1500;"SCR"
OUTPUT @B1500;"SCR 5"