Agilent B1500A/B1505A Programming Guide, Edition 11 4-41
Command Reference
AITM
Example
Statements
OUTPUT @B1500;"AIT 2,3,.001"
AITM
This command is valid for the current measurement by HRSMU. This command
sets the operation mode of the high-resolution ADC that is set to the power line
cycle (PLC) mode by the AIT 1, 2, N command.
The mode setting is cleared by the *RST or a device clear (HP BASIC CLEAR)
command.
Syntax AITM mode
Parameters mode : Operation mode. Integer expression. 0 or 1.
0: B1500 standard operation mode. Initial setting.
1: Classic operation mode. This is similar mode as Agilent 4156’s PLC
mode.
AITM?
This command returns the operation mode of the high-resolution ADC that is set by
the AITM command.
The mode setting is cleared by the *RST or a device clear (HP BASIC CLEAR)
command.
Syntax AITM?
Query Response mode<^EOI>
Operation mode of the high-resolution ADC. 0 (standard) or 1 (classic).
ALS
This command sets the ALWG sequence data.
To generate an arbitrary linear waveform, the SPGU channels need both sequence
data and pattern data. The sequence data must be used by and in common with all
SPGU channels installed in the B1500. And the pattern data must be set to each
SPGU channel.
See “SPGU Module” on page 2-53 for details on the ALWG output and the data
format.