Agilent Technologies B1500A Servisní příručka Strana 553

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 552
Agilent B1500A/B1505A Programming Guide, Edition 11 4-237
Command Reference
XE
If any channel is set to the high voltage state (forcing more than the allowable
voltage for the interlock open condition, or voltage compliance set to more than
it) after the trigger (XE), the interlock terminal must be shorted.
The commands shown in Table 4-33 must be entered before the XE command.
Table 4-33 Required Commands before XE, by Measurement Mode
Measurement Mode Commands
Spot CN, MM, DV or DI
Staircase sweep CN, MM, WV or WI
Pulsed spot CN, MM, PT, PV or PI
Pulsed sweep CN, MM, PT, PWV or PWI
Staircase sweep with pulsed bias CN, MM, PT, WV or WI, PV or PI
Quasi-pulsed spot CN, MM, BDV
Sampling CN, MM, MCC, ML, MT, MSC, MI, MV,
MSP
Quasi-static CV CN, MM, QST, QSV
Liner search CN, MM, LSV or LSI, LGV or LGI
Binary search CN, MM, BSV or BSI, BGV or BGI
Multi channel sweep CN, MM, WI or WV, WNX
Spot C CN, MM, IMP, FC, ACV, DCV
CV (DC bias) sweep CN, MM, IMP, FC, ACV, WDCV
CV (AC level) sweep CN, MM, IMP, FC, DCV, WACV
C-f sweep CN, MM, IMP, DCV, ACV, WFC
Pulsed spot C CN, MM, IMP, FC, ACV, PTDCV, PDCV
Pulsed sweep CV CN, MM, IMP, FC, ACV, PTDCV, PWDCV
C-t sampling CN, MM, IMP, FC, ACV, MDCV, MTDCV
Multi channel pulsed spot CN, MM, MCPT, MCPNT, MCPNX
Multi channel pulsed sweep CN, MM, MCPT, MCPNT, MCPWS,
MCPWNX
Zobrazit stránku 552
1 2 ... 548 549 550 551 552 553 554 555 556 557 558 ... 597 598

Komentáře k této Příručce

Žádné komentáře