4-136 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ML
Vcomp : Voltage compliance value (in V). Numeric expression. See Table 4-6 on
page 4-23, Table 4-8 on page 4-25, or Table 4-11 on page 4-27 for each
measurement resource type. If you do not specify this parameter,
Vcomp is set to the previous setting.
Example
Statements
OUTPUT @B1500;"MI 1,18,0,5E-5,10"
See Also “MT”, “MCC”, “MSC”
ML
The ML command sets the sampling mode, linear or logarithmic. For the
logarithmic sampling, this command also specify the number of measurement data
to be returned.
If you do not execute this command, the last sampling mode is effective.
Initialization such as the *RST command sets the linear sampling mode.
Syntax ML mode
Parameters mode Sampling mode, linear or logarithm.
1: linear sampling, initial setting.
2: logarithmic sampling, 10 data/decade.
3: logarithmic sampling, 25 data/decade.
4: logarithmic sampling, 50 data/decade.
5: logarithmic sampling, 100 data/decade.
6: logarithmic sampling, 250 data/decade.
7: logarithmic sampling, 500 data/decade.
Example
Statements
OUTPUT @B1500;"ML 2"
MM
The MM command specifies the measurement mode and the channels used for
measurements. This command must be entered to specify the measurement mode.
For the high speed spot measurements, do not enter the MM command.
Syntax mode= 1, 2, 10, 16, 18, 27, or 28: