4-68 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
CORRST
OUTPUT @B1500;"SPV 101,1,-0.5,0.5"
OUTPUT @B1500;"CORRSER? 101,1,1E-7,1E-8,10"
ENTER @B1500;A,B
In this example, the voltage measurement data is returned to the variable B, and the
load impedance calculation data is returned to the variable A.
See Also “SER”, “SER?”
NOTE Terminal voltage measurement and load impedance calculation
SPGU performs voltage measurement and impedance calculation by executing the
CORRSER? command. Followings are the recommended measurement conditions
for the voltage measurement.
Output voltage: > 1 V
Minimum load impedance: 40 Ω
Maximum load impedance: 500 Ω (1 V), 2000 Ω (2 V), 5000 Ω (10 V)
Figure 4-3 Voltage Monitor and CORRSER? Setup Parameters
CORRST
This command enables or disables the MFCMU open/short/load correction function.
Before setting a function to ON, perform the corresponding correction data
measurement by using the CORR? command.
delay interval
time
Ch1 output
Ch2 output
Start measurement
Meas[1]
Meas[2]
Meas[n]
delay + int erval x n
Measurement value = average(Meas[n] )
1 pulse width or 1 ALWG sequence
Pulse period >
n: count