Agilent B1500A/B1505A Programming Guide, Edition 11 4-153
Command Reference
PTDCV
UHCU: 10 μs to 1 ms and duty ratio maximum 0.4 % for 500 A range, 10
μs to 500 μs and
duty ratio maximum 0.1 % for 2000 A range, 2 μs
resolution.
UHVU: 100 μs to 1 ms for 100 mA range, 100 μs to 2 s for other range,
2 μs resolution.
HVMCU: 10 μs to 1 ms for 100 mA range, 10 μs to 100 μs for 1 A/2 A
range, 2 μs resolution.
period : Pulse period (in seconds). Numeric expression. 0, -1, or 5 ms to 5.0 s.
0.1 ms resolution. For using UHVU, minimum pulse period is 10 ms.
Initial setting = 10 ms. Default setting = 0.
period width + 2 ms (for width 100 ms)
period width + 10 ms (for 100 ms < width)
period =-1: Automatically set to the effective minimum period.
period =0: Automatically set to the longest one of the followings.
Minimum period given by the pulse width and the duty ratio
Pulse period = 5 ms (for width 3 ms)
Pulse period = width + 2 ms (for 3 ms < width 100 ms)
Pulse period = width + 10 ms (for 100 ms < width)
Tdelay : Trigger output delay time (in seconds). Numeric expression. 0 to width.
0.1 ms resolution. Initial or default setting = 0.
This parameter is the time from pulse leading edge to timing of trigger
output from a trigger output terminal.
PTDCV
This command sets the hold time, pulse width, pulse period, and trigger output delay
time of the pulsed spot C measurement (MM19) or pulsed sweep CV measurement
(MM20).
Syntax For pulsed spot C measurement:
PTDCV hold,width[,period[,Tdelay]]
For pulsed sweep CV measurement:
PTDCV hold,width,period[,Tdelay]