4-158 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
PWV
PWV
The PWV command specifies the pulsed voltage sweep source and its parameters.
This command also clears the settings of the PWI, WSV and WSI commands. The
settings specified by this command are cleared by the PWI command.
To set the timing of output pulse and measurement, use the PT and AIT2 commands.
Syntax PWV chnum,mode,range,base,start,stop,step[,Icomp[,Pcomp]
]
Parameters chnum : SMU pulsed sweep source channel number. Integer expression. 1 to 10
or 101 to 1001. See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression. 1 to 4.
1: Linear sweep (single stair, start to stop.)
2: Log sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
4: Log sweep (double stair, start to stop to start.)
range: Ranging type for pulsed voltage sweep. Integer expression. The output
range will be set to the minimum range that covers base, start, and stop
values. For the limited auto ranging, the instrument never uses the
range less than the specified range. See Table 4-4 on page 4-21.
base, start,
stop : Pulse base, start, or stop voltage (in V). Numeric expression. See Table
4-7 on page 4-24, Table 4-9 on page 4-26, Table 4-10 on page 4-26,
Table 4-12 on page 4-27, Table 4-14 on page 4-28, or Table 4-15 on
page 4-28 for each measurement resource type. If the output voltage is
greater than the allowable voltage for the interlock open condition, the
interlock circuit must be shorted. For the log sweep or using 3000 V
range of HVSMU, base, start, and stop must have the same polarity.
For using UHVU, base, start, and stop must have the same polarity.
step : Number of steps for pulsed sweep. Numeric expression. 1 to 10001.
Icomp : Current compliance (in A). Numeric expression. See Table 4-7 on page
4-24, Table 4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on
page 4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for
each measurement resource type. If you do not specify Icomp, the
previous value is set.