
2-76 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Trigger Function
PA/PAX/WS/WSX
Commands
The commands put the B1500 in the trigger wait state. The B1500 can recover from
the wait state if an external trigger is sent to a trigger input port. You can use the
commands regardless of the trigger type.
If you use the PA or PAX command to put the B1500 in the trigger wait state, send
the TM3 command before the PA or PAX command.
Table 2-7 Type of Trigger Input
Type B1500 Operation by Input Trigger
Command
a
a. t selects trigger input terminal, Ext Trig In or a digital I/O path.
p selects positive or negative logic of the trigger.
m selects Case 1 or Case 2 of the trigger type 2 (see Figure 2-47).
1 Starts the measurement specified by the MM command. TGP t,1,p,1
TM3
2 Available for the
staircase sweep, multi channel sweep,
pulsed spot, pulsed sweep, staircase sweep with pulsed bias,
multi channel pulsed spot, multi channel pulsed sweep, and
CV(DC)/CV(AC)/C-f sweep measurements.
The sweep source starts to set the sweep step output.
The pulse source starts the pulse output at 110 μs after
receiving the trigger.
TGP t,1,p,2
TGSI m
3 Available for the staircase sweep, multi channel sweep,
and
CV(DC)/CV(AC)/C-f sweep measurements.
Waits for the measurement trigger delay, and starts the
sweep step measurement.
TGP t,1,p,3
Komentáře k této Příručce