4-138 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
MM
Table 4-28 Measurement Mode
mode
Measurement
mode
Related source setup
command
SMU
MF
CMU
UHCU
a
HV
MCU
b
UHVU
c
HP/
MP/
HR
MC/
HC/
HV
1 Spot DI, D V Yes Yes Yes
2 Staircase sweep WI, WV, WT, WM,
WSI, WSV
Yes Yes Yes
3 Pulsed spot PI, PV, PT Yes Yes Yes Yes Yes
4 Pulsed sweep PWI, PWV, PT, WM,
WSI, WSV
Yes Yes Yes Yes Yes
5 Staircase sweep with
pulsed bias
WI, WV, WM, WSI,
WSV, PI, PV, PT
Yes Yes Yes Yes Yes
9 Quasi-pulsed spot BDV, BDT, BDM Yes
10 Sampling MCC, MSC, ML, MT,
MI, MV
Yes Yes Yes
13 Quasi-static CV QSV, QST, QSM Yes
14 Linear search LSV, LSI, LGV, LGI,
LSM, LSTM, LSSV,
LSSI, LSVM
Yes Yes Yes
15 Binary search BSV, BSI, BGV, BGI,
BSM, BST, BSSV,
BSSI, BSVM
Yes Yes Yes
16 Multi channel sweep WI, WV, WT, WM,
WNX
Yes Yes Yes
17 Spot C FC, ACV, DCV Yes
18 CV (DC bias) sweep FC, ACV, WDCV,
WMDCV, WTDCV
Yes Yes Ye s
19 Pulsed spot C PDCV, PTDCV Yes
20 Pulsed sweep CV PWDCV, PTDCV Yes
22 C-f sweep WFC, ACV, DCV,
WMFC, WTFC
Yes
23 CV (AC level)
sweep
FC, WACV, DCV,
WMACV, WTACV
Yes
26 C-t sampling MSC, MDCV, MTDCV Yes
27 Multi channel pulsed
spot
MCPT, MCPNT,
MCPNX
Yes Yes Yes Yes Yes
28 Multi channel pulsed
sweep
MCPT, MCPNT,
MCPNX, MCPWS,
MCPWNX, WNX
Yes Yes Yes Yes Yes
Komentáře k této Příručce