Agilent Technologies B1500A Servisní příručka Strana 386

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 385
4-70 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
DCORR
DCORR
This command disables the MFCMU open/short/load correction function and sets
the open/short/load standard calibration value or reference value to the B1500. This
command also clears the correction data.
The reference values set by this command are cleared by turning off power.
Syntax DCORR chnum
,corr,mode,primary,secondary
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
corr : Correction mode. Integer expression. 1, 2, or 3.
1: Open correction
2: Short correction
3: Load correction
mode : Measurement mode. Integer expression. 100 or 400.
100: Cp-G (for open correction)
400: Ls-Rs (for short or load correction)
primary : Primary reference value of the standard. Numeric expression.
Cp value for the open standard. in F.
Ls value for the short or load standard. in H.
secondary : Secondary reference value of the standard. Numeric expression.
G value for the open standard. in S.
Rs value for the short or load standard. in Ω.
Example
Statements
OUTPUT @B1500;"DCORR 9,3,400,0.00001,49.8765"
DCORR?
This command returns the calibration value or the reference value of the standard
used for the MFCMU open/short/load correction.
Syntax DCORR? chnum
,corr
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
Zobrazit stránku 385
1 2 ... 381 382 383 384 385 386 387 388 389 390 391 ... 597 598

Komentáře k této Příručce

Žádné komentáře