2-92 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Initial Settings
Table 2-13 Initial Settings of Mainframe, SMU, and CMU
Setup Item Initial Setting Commands
Auto calibration off CM
SMU output switch open CN, CL
SMU filter/series resistor off/off FL/SSR
ASU path/1 pA auto range/indicator SMU side/disable/enable SAP/SAR/SAL
SCUU path/indicator open/enable SSP/SSL
SMU current measurement range with pulse compliance range RI
without pulse auto
SMU voltage measurement range with pulse compliance range RV
without pulse auto
SMU A/D converter high speed ADC AAD
SMU ADC Integration time high speed ADC: auto, non parallel AIT, PAD
high resolution ADC: auto AIT
SMU ADC zero function off AZ
SMU AV command parameter number=1, mode=0 AV
CMU measurement parameter Cp-G IMP
CMU measurement range auto RC
CMU ADC integration time auto ACT
CMU correction/compensation Open/Short/Load: off/off/off, Phase compensation: auto OPEN/SHOR/LOAD, ADJ
CMU AC signal 0 V, 1 kHz ACV
Sweep source parameters cleared WV, WSV, WI, WSI, WDCV
Pulse source parameters cleared PV, PI
Pulse sweep source parameters cleared PWV, PWI
Search source parameters cleared BSV, BSSV, BSI, BSSI, LSV, LSSV, LSI, LSSV
Search monitor parameters cleared BGV, BGI, LGV, LGI
Search measurement data source output value only BSVM, LSVM
Quasi-pulse source parameters cleared BDV
Quasi-pulsed spot measurement mode voltage BDM
Quasi-pulse settling detection interval short BDM
Sampling source cleared MI, MV
Sampling interval, sampling point 2 ms, 1000 points MT
Automatic abort function off WM, BSM, LSM, WMDCV, MSC
Output after measurement start value (bias value for MSC) WM, BSM, LSM, WMDCV, MSC
Pulse width 0.001 s PT
Pulse period 0.01 s PT
Hold time 0 s WT, PT, BDT, BST, LSTM, WTDCV, MT
Delay time 0 s WT, PT, BDT, BST, LSTM, WTDCV
Step delay time 0 s WT, WTDCV
Trigger delay time 0 s WT, PT, WTDCV
Trigger mode XE, TV, TI, or GET TM
Trigger port Ext Trig In Start Measurement trigger input TGP
Ext Trig Out Measurement Completion trigger output TGP
Digital I/O cleared TGP
Trigger condition of
Start Step Output Setup trigger
with trigger for first sweep step TGSI
Type of output trigger edge trigger TGXO, TGSO, TGMO
Digital I/O port output for all port ERM
Program memory cleared. Not cleared by *RST command or device clear. SCR
Value of internal variable (%In, %Rn) 0 VAR
Data output format ASCII with header, CR/LF^EOI FMT
Data output buffer cleared BC
Status byte Only bit 6 is enabled. *SRE
Error code register cleared ERRX?, ERR?