Agilent Technologies B1500A Servisní příručka Strana 444

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 443
4-128 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
MCPNT
Example
Statements
OUTPUT @B1500;"MCC"
OUTPUT @B1500;"MCC 1,2,3"
MCPNT
The MCPNT command sets the delay time and the pulse width of the pulse output
channels. This command is effective for the multi channel pulsed spot or sweep
measurement set by MM 27 or MM 28.
Syntax MCPNT chnum,delay,width
Parameters chnum : SMU pulsed source channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
delay : Delay time (in seconds) from the beginning of the pulse period to the
beginning of the transition from base to peak. Numeric expression.
Effective value depends on the measurement resource type. See Table
4-27 on page 4-128.
width : Pulse width (in seconds). Numeric expression. Effective value depends
on the measurement resource type. See Table 4-27 on page 4-128.
Table 4-27 Effective value for delay and width
HR/MP/HP
SMU
HCSMU /
DHCSMU
MCSMU HVSMU UHCU UHVU HVMCU
delay 00 to period-width
period is set by MCPT command.
Resolution: 2 μs, Initial value: 0
width 500 μs to 2 s
Resolution:
100 μs
Initial
setting: 1 ms
Same value
for all
HR/MP/HPS
MU.
50 μs to 1
ms, duty
ratio
maximum 1
%, for 20 A
range and 40
A range
50 μs to 2 s
for
other range
Resolution:
2 μs
Initial
setting: 1 ms
10 μs to
100 ms,
duty ratio
maximum
5 %, for 1
A range
10 μs to 2 s
for
other range
Resolution
: 2 μs
Initial
setting: 1
ms
500 μs to 2
s
Resolution
: 2 μs
Initial
setting: 1
ms
10 μs to 1 ms,
duty ratio
maximum 0.4
%, for 500 A
range
10 μs to 500
μs, duty ratio
maximum 0.1
%, for 2000
Arange
Resolution:
2 μs
Initial setting:
500 μs
100 μs to 1
ms for 100
mA range
100 μs to 2
s for other
range
Resolution:
2 μs
Initial
setting: 1 ms
10 μs to 1
ms for 100
mA range
10 μs to
100 μs for
1 A/2 A
range
Resolution:
2 μs
Initial
setting: 100
μs
Zobrazit stránku 443
1 2 ... 439 440 441 442 443 444 445 446 447 448 449 ... 597 598

Komentáře k této Příručce

Žádné komentáře