Agilent B1500A/B1505A Programming Guide, Edition 11 4-23
Command Reference
Table 4-6 HR/MP/HPSMU Current Source Setup Parameters
1
Output range
(actually used)
a
a. Output ranges 1 pA to 1 nA are available only for DC output, not for pulse output.
Setting
resolution in A
current, start, stop,
base, bias, or pulse
in A
Maximum Vcomp value in V
HRSMU MPSMU HPSMU
1 pA 1E-15 0 to ± 1.15 E-12 ±100 NA NA
10 pA 5E-15 0 to ± 11.5 E-12 ±100
100 pA 5E-15 0 to ± 115 E-12 ±100
1 nA 50E-15 0 to ± 1.15 E-9 ±100 ±100 ±200
10 nA 500E-15 0 to ± 11.5 E-9 ±100 ±100 ±200
100 nA 5E-12 0 to ± 115 E-9 ±100 ±100 ±200
1 μA 50E-12 0 to ± 1.15E-6 ±100 ±100 ±200
10 μA 500E-12 0 to ± 11.5E-6 ±100 ±100 ±200
100 μA 5E-9 0 to ± 115E-6 ±100 ±100 ±200
1 mA 50E-9 0 to ± 1.15E-3 ±100 ±100 ±200
10 mA 500E-9 0 to ± 11.5E-3 ±100 ±100 ±200
100 mA 5E-6 0 to ± 20E-3 ±100 ±100 ±200
to ± 50E-3 ±40 ±40 ±200
to ± 100E-3 ±20 ±20 ±100
to ± 115E-3 NA NA ±100
1 A 50E-6 0 to ± 50E-3 ±200
to ± 125E-3 ±100
to ± 500E-3 ±40
to ± 1 ±20
1. The following commands are used for setting the MP/HR/HPSMU to the current source.
DI, TDI, WI, WSI, WNX, PI, PWI, LSI, LSSI, BSI, BSSI, MI, MCPNX, MCPWNX