2-88 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Initial Settings
Table 2-10 SMU Settings
Setup Item Initial Setting Commands
Output switch open CN, CL
Filter off FL
Series resistor off SSR
ASU path/1 pA auto range/indicator SMU side/disable/enable SAP/SAR/SAL
Current measurement range with pulse compliance range RI
without pulse auto
Voltage measurement range with pulse compliance range RV
without pulse auto
A/D converter high speed ADC AAD
ADC integration time high speed ADC: auto, non parallel AIT, PAD
high resolution ADC: auto AIT
ADC zero function off AZ
AV command parameter number=1, mode=0 AV
Sweep source parameters cleared WV, WSV, WI, WSI
Pulse source parameters cleared PV, PI
Pulse sweep source parameters cleared PWV, PWI
Search source parameters cleared BSV, BSSV, BSI,
BSSI, LSV, LSSV,
LSI, LSSV
Search monitor parameters cleared BGV, BGI, LGV,
LGI
Search measurement data source output value only BSVM, LSVM
Quasi-pulse source parameters cleared BDV
Quasi-pulsed spot measurement mode voltage BDM
Komentáře k této Příručce