4-142 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
MT
For the linear sampling: 100001 / (number of measurement channels)
For the log sampling: 1 + (number of data for 11 decades)
h_base Hold time of the base value output until the bias value output. Numeric
expression. in seconds. 0 (initial setting) to 655.35 s, resolution 0.01 s.
Sampling
Operation
Sampling measurement will be started by a measurement trigger such as the XE
command or an external trigger, and performed as shown below. Before the
measurement trigger, the source channels set by the DI/DV commands will start
output at the timing of the DI/DV command execution.
1. By the measurement trigger, source channels set by the MI/MV commands start
the base value output. Each source channel controls the output simultaneously.
2. h_base seconds later, the source channels change the output to the bias value.
The channels keep the value until the end of the sampling measurement.
3. Another h_bias seconds later, the measurement channels start measurement for
the first sampling point. The measurement channels perform the measurement in
series by the order set to the MM command.
4. After that, the following operation is repeated with the specified time interval.
Measurement channels start measurement if they are ready to measure.
Measurement channels keep the condition if they are busy.
This operation is repeated until the number of measurement result data reaches
to the specified number of measurement data.
For the linear sampling with interval < 2 ms, if the total measurement time runs
over the specified time interval × number, the sampling measurement will be
stopped even if the number of measurement result data is less than the specified
number.
For the log sampling, the B1500 holds only the data that can be plotted on the
log scale in the same distance as close as possible. Only the held data is counted
in the number of measurement result data.
5. The sampling measurement is completed. And the source channel set by the
MI/MV command forces the base or bias value specified by the MSC command.
The source channel set by the DI/DV command keeps its output.
The index data (max. 9999999) and the time data returned with the measurement
data will be as shown in the following formula. However, long measurement or busy
status may cause unexpected time data.
time data = t + h_bias + (index data -1) × interval