Agilent Technologies B1500A Servisní příručka Strana 411

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 410
Agilent B1500A/B1505A Programming Guide, Edition 11 4-95
Command Reference
ERPFGA?
Pulse: 1 A, maximum duty ratio is 5 %.
Example
Statements
OUTPUT @B1500;"ERPFGA 5"
ERPFGA?
This query command returns the slot number for the MC/HCSMU module
connected to the Gate input of Agilent N1265A ultra high current expander/fixture.
Syntax ERPFGA?
Query Response gsmu<CR/LF^EOI>
gsmu : Slot number for the MC/HCSMU connected to the Gate input.
Example
Statements
OUTPUT @B1500;"ERPFGA?"
ENTER @B1500;gsmuUHC
ERPFGP
This command sets the connection status for the MC/HCSMU output connected to
the Gate input of Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFGP state
Parameters state : Connection status. Integer expression. 0 or 1.
0: Open, initial setting.
1: Connected.
Example
Statements
OUTPUT @B1500;"ERPFGP 1"
ERPFGP?
This query command returns the connection status for the MC/HCSMU module
connected to the Gate input of Agilent N1265A ultra high current expander/fixture.
Syntax ERPFGP?
Zobrazit stránku 410
1 2 ... 406 407 408 409 410 411 412 413 414 415 416 ... 597 598

Komentáře k této Příručce

Žádné komentáře