Agilent B1500A/B1505A Programming Guide, Edition 11 4-83
Command Reference
ERHPP?
0: Open, no module is connected, initial setting
1: Connects to HVSMU port
2: Connects to HCSMU port
3: Connects to HPSMU port
4: Connects to HVSMU port, also connects the series resistor
Example
Statements
OUTPUT @B1500;"ERHPP 3"
ERHPP?
The ERHPP? command returns the input-to-output connection path of Agilent
N1258A/N1259A module selector.
Syntax ERHPP?
Query Response path<CR/LF^EOI>
Input-to-output connection path. Integer expression. 0, 1, 2, 3, or 4.
0: Open, no module is connected
1: Connects to HVSMU port
2: Connects to HCSMU port
3: Connects to HPSMU port
4: Connects to HVSMU port, also connects the series resistor
Example
Statements
OUTPUT @B1500;"ERHPP?"
ENTER @B1500;A
ERHPR
Only for the N1258A module selector users. This command sets the logical state ON
or OFF to the specified relay control output pin. See “External Relay Control
Output” on page 2-65 for the external relay control function.
Execution
Conditions
Digital I/O port must be set to the N1258A/N1259A control mode using the
ERMOD 2 command.
External relay control function must be enabled using the ERHPE 1 command.
Syntax ERHPR pin,state