2-4 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
Spot Measurements
Spot measurement is performed as shown below. The measurement channel
performs one point measurement.
Figure 2-1 Spot Measurements
1. The source channel starts output by the DV or DI command. Multiple channels
can be set.
2. The measurement channel starts measurement by a trigger, such as the XE
command. If the trigger is received during the settling time of the source
channels, measurement starts after the settling time.
If you use multiple measurement channels, the channels perform measurement
in the order defined in the MM command. Parallel measurement is also
available, see “Parallel Measurement Function” on page 2-48.
3. After measurement, the source channels continue the source output.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE The DV command is used to force voltage, and the DI command is used to force
current.
Voltage or current
DV/DI DV/DI Measurement trigger (e.g. XE)
Measurement time
Channel 1 output
Channel 2 output
Setup value
Previous value
Setup value
Previous value
Voltage or current
Time
: Measurement