Agilent B1500A/B1505A Programming Guide, Edition 11 2-89
Remote Mode Functions
Initial Settings
Quasi-pulse settling detection interval short BDM
Sampling source cleared MI, MV
Sampling interval, sampling point 2 ms, 1000 points MT
Automatic abort function off WM, BSM, LSM,
MSC
Output after measurement start value (bias value for MSC)
Hold time 0 s WT, PT, BDT, BST,
LSTM
Delay time 0 s
Step delay time 0 s WT
Trigger delay time 0 s WT, PT
Pulse width 0.001 s PT
Pulse period 0.01 s PT
Setup Item Initial Setting Commands