Agilent B1500A/B1505A Programming Guide, Edition 11 2-7
Remote Mode Functions
Measurement Modes
6. After the pulse width, the pulse source channels apply the pulse base value, and
keep it.
If the next trigger occurs within the pulse period, pulse output is as follows if the
trigger interval is longer than several 10 ms.
If the rest of the pulse period is longer than the hold time as shown in Figure
2-3, the pulse source waits for the rest, then starts the pulse output.
If the rest of the pulse period is shorter than the hold time, the pulse source
waits for the hold time since the last trigger, then starts the pulse output.
If the hold time, pulse period, and trigger interval are very short, pulse settling
time may be shown.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE The MCPNT command sets the delay time and pulse width. The MCPNX command
sets the pulse output. The base and peak values must have the same polarity for the
current pulse.