Agilent B1500A/B1505A Programming Guide, Edition 11 4-169
Command Reference
RU
• Program memory setup data
• Self-calibration data
• MFCMU phase compensation data
• MFCMU open/short/load correction data
Example
Statement
OUTPUT @B1500;"*RST"
RU
The RU command sequentially executes the internal memory programs.
Execution
Conditions
The specified programs have been stored by using the ST and END commands,
from the start program number through the stop program number.
Syntax RU start,stop
Parameters start : Start program number. Numeric expression. 1 to 2000.
stop : Stop program number. Numeric expression. 1 to 2000.
where stop value must be greater than or equal to the start value.
Example
Statements
OUTPUT @B1500;"RU 1,10"
OUTPUT @B1500;"RU 3,6"
RV
The RV command specifies the voltage measurement range or ranging type. In the
initial setting, the auto ranging is set. The range changing occurs immediately after
the trigger (that is, during the measurements). Voltage measurement channel can be
decided by the CMM command setting and the channel output mode (voltage or
current).
For the high speed spot measurement, use the TV/TTV command.
The range setting is cleared by the CL, CA, IN, *TST?, *RST or a device clear (HP
BASIC CLEAR) command.
Syntax RV chnum,range
Parameters chnum : SMU voltage measurement channel number. Integer expression. 1 to 10
or 101 to 1001. See Table 4-1 on page 4-16.