4-190 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
STGP?
STGP?
This query command returns the trigger output state of the specified SPGU channel.
Syntax STGP? chnum
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
Query Response state<CR/LF^EOI>
0 Trigger output disabled.
1 Output trigger signals synchronized to the pulses (PG mode), or to the start of
the ALWG sequence.
2 Output a trigger when the ALWG pattern changes, or at start of the first pattern.
3 Output triggers at the start of every ALWG pattern.
Example
Statements
OUTPUT @B1500;"STGP? 101"
ENTER @B1500;A
TACV
This command forces AC voltage from the MFCMU, and returns the time data from
timer reset to the start of output.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
The FC command has been executed to set the frequency of the AC voltage.
This command is not effective for the 4 byte binary data output format (FMT3 and
FMT4).
Syntax TACV chnum,voltage
Parameters chnum : MFCMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
voltage : Oscillator level of the output AC voltage (in V). Numeric expression.
0 (initial setting) to 0.25 V, 0.001 V step.
Remarks To read the time data with the best resolution (100 μs), the timer must be cleared
every 100 sec or less for the FMT1, 2, or 5 data output format, or
every 1000 sec or less for the FMT 11, 12, 15, 21, 22, or 25 data output format.