Agilent Technologies B1500A Servisní příručka Strana 85

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 598
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 84
Agilent B1500A/B1505A Programming Guide, Edition 11 1-67
Programming Basics
Programming Tips
To Use Programs for Agilent E5260/E5270
Agilent B1500 supports most of the commands and the data output format supported
by the Agilent E5260/E5270 Series of Parametric Measurement Solutions. To reuse
the programs created for the Agilent E5260/E5270, confirm the following and
modify the programs if necessary.
To remove all unsupported commands
Some commands are not supported owing to differences in the mainframe. See
Table 1-19 that shows the commands not supported by the B1500. The
commands will not cause errors because the B1500 will ignore these commands.
However remove these commands to reduce the load.
*CAL?, RCV, *TST? command
Use 11 to specify the B1500 mainframe instead of 9 that indicates the
E5260/E5270 mainframe.
DIAG? command
The B1500 does not support the front panel key test and the beeper test. So, do
not use the parameter item=2 and 5.
Table 1-19 E5260/E5270 FLEX Commands Unsupported
Category Command
Display Control RED, DFM, SPA, MPA, SCH, MCH
Keyboard Control KLC
Zobrazit stránku 84