Agilent B1500A/B1505A Programming Guide, Edition 11 4-223
Command Reference
WNX
measurement. There is no restrictions for the output mode (voltage or current) of the
sweep sources. This command is available for the multi channel sweep
measurement (MM16 and MM28).
For MM16, the primary sweep source is set by the WI or WV command.
For MM28, the primary sweep source is set by the WNX or MCPWNX command
defined with the parameter N=1.
Sweep mode, linear or log, and the number of sweep steps are set by the WI or WV
command for MM16, or the MCPWS command for MM28.
This command setting is cleared by the WI, WV, or WNCC command.
Execution
Conditions
For MM16, the WI or WV command must be entered before the WNX command.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax WNX N,chnum,mode,range,start,stop[,comp[,Pcomp]]
Parameters N : Source number. Integer expression. 2 to 10 for MM16. 1 to 10 for
MM28. See Remarks below.
chnum : SMU sweep source channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
mode : Sweep source type. Integer expression. 1 or 2.
1: Voltage sweep source
2: Current sweep source
range : Ranging type for synchronous sweep output. Integer expression.
For voltage source (mode=1): See Table 4-4 on page 4-21.
The B1500 usually uses the minimum range that covers both start
and stop values to force the staircase sweep voltage. However, if
you set Pcomp and if the following formulas are true, the B1500
changes the output range dynamically (20 V range or above). Range
changing may cause 0 V output in a moment. For the limited auto
ranging, the instrument never uses the range less than the specified
range.
comp > maximum current for the output range
Pcomp/output value > maximum current for the output range
For current source (mode=2): See Table 4-5 on page 4-22.